Johnson-level-tool-mfg 40-6580 Manual de usuario Pagina 22

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 30
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 21
22 ©2008 Johnson Level & Tool
8. Self-Check and Calibration
IMPORTANT: It is the responsibility of the user to verify the
calibration of the instrument before each use.
Accuracy Check
1. Put the laser on a tripod 50’ away from the wall. Put the laser
on the tripod with X+ axis towards wall.
2. Power on and after self-leveling,
using the detector, make a
mark “A on the wall where the detector indicates on grade with
the laser beam.
3. Turning the instrument 90º, Y+, X-, Y-, after the laser self-levels,
mark “B” for Y+, “C” for X- and “D” for Y- on the front wall. Make
sure points B, C, D are in the same vertical line as point A.
4. Measure the vertical distance between the highest and lowest
points between A, B, C, D and mark that “h”.
5. If “h” is less than 1/32”, the accuracy is good. If it is more than
1/32”, the accuracy is beyond its tolerance and the laser needs to
be recalibrated.
50’
2406H 5/28/09 9:04 AM Page 22